Digital non-glass pH sensor
Ceramax CPS341D

Ceramax CPS341D - Memosens enamel sensor for tough process conditions ©Endress+Hauser
Ceramax CPS341D - Unbreakable, perfectly suited for food applications ©Endress+Hauser
Ceramax CPS341D - Long-term stable over many years ©Endress+Hauser
Ceramax CPS341D - Detail view of pH-sensitive enamel coating ©Endress+Hauser

The pH-value of a liquid can be calculated using the potentiometric measurement principle. This video shows what it is about and how this measuring principle works.

Memosens converts the measured value to a digital signal and transfers it to the transmitter without a contacting connection. This video shows how Memosens optimizes process control and maintenance.

Memosens enamel electrode for food and life sciences industries

Ceramax CPS341D is the long runner for hygienic applications in food and life sciences industries. The sensor guarantees long-term stability and lowest maintenance wherever it is applied. Thanks to Memosens digital sensor technology, CPS341D combines maximum process and data integrity with simple operation. It resists corrosion and moisture, enables lab calibration and facilitates predictive maintenance.

  • Benefits

    • Long-term stability over many years

    • Hygienic design: Suitable for cleaning in place (CIP) and sterilization in place (SIP)

    • High mechanical stability due to pH-sensitive enamel on a steel carrier

    • Insensitive to soiling thanks to large pH measuring surface

    • ...

  • Field of application

    • Food production, also highly pasty media

    • Beverage production and filling

    • Life sciences:
      - Water treatment
      - Active ingredient production
      - Active ingredient preparation
      - Biotechnology

Features and specifications

Back

Details

  • Measuring principle

    Glass electrode

  • Application

    Hygienic and sterile applications, nonglass, shatterproof. Pharmaceuticals, biotechnology, vontaminating media

  • Characteristic

    pH sensitive enamel, CIP- and SIP-able, liquid KCl-filled reference

  • Measurement range

    pH -2 - 14

  • Measuring principle

    Ground diaphragm, liquid KCl-filled reference

  • Design

    Only one shaft length, different hygienic process connectors

  • Material

    pH sensitive enamel on a steel carrier

  • Dimension

    Diameter 12 mm (0.47 inch), shaft length 120 mm (4.72 inch)withour process connector

  • Process temperature

    0 ... 140°C, 32 - 280°F

  • Process pressure

    max. 6 bar (0 - 87 psi)

  • Temperature sensor

    NTC30k

  • Connection

    Contactless, inductive, digital connection head with Memosens technology

Documents / Manuals / Software

Accessories / Spare parts

Nr
Order code
Specification