Specs at a glance
-
Accuracy
0.075% of individual sensor,
"PLATINUM" 0.05% of individual sensor -
Process temperature
–25...+150°C
(–13...+302°F) -
Pressure measuring range
100mbar...40bar
(1.5psi...600psi) -
Process pressure / max. overpressure limit
60 bar (900 psi)
-
Material process membrane
Ceramic
316L, AlloyC -
Measuring cell
100 mbar...40 bar
(1.5 psi...600 psi)
Field of application
The electronic dp Deltabar FMD71 is a differential pressure system, used to measure the pressure or level, volume or mass of liquids in pressurized tanks or distillation columns/evaporators. The high pressure sensor (HP) measures the hydrostatic pressure. The low pressure sensor (LP) measures the head pressure. The level is calculated in the transmitter using these two digital values. The electronic dp system eliminates issues of traditional differential pressure measurements.
For differential pressure, level, volume or mass measurement of liquids in pressurized tanks and distillation columns/evaporators.
Process connections: Threads, flanges and hygienic connections
Process temperature: -40 up to +150°C (-40 to +302°F)
Measuring range: 100mbar to +40bar (6 to 600psi)
Accuracy: Single sensor up to ±0.05%, system up to ±0.07%
International explosion protection certificates, hygienic approvals
Benefits
Eliminates traditional mechanical issues resulting in greater process availability and reliability
Overload-resistant high purity ceramic sensor (99.9% Al2O3)
Safety risks are minimized with the new electronic differential pressure system architecture and design
Lowest total cost of ownership due to reduced installation time, maintenance, downtime and spare requirements
Multivariable level measurement: HART-based differential pressure, head pressure and sensor temperatures from one system
Continuous health indication of the entire system via HART-based diagnostic
High reproducibility and long-term stability
Videos
-
Pressure measuring principle
-
The new electronic dp system
Extended selection
-
High-end products
- Highly functional and convenient
Technical excellence
Simplicity

-
Standard products
- Reliable, robust and low-maintenance
Technical excellence
Simplicity
-
Specialized products
- Designed for demanding applications
Technical excellence
Simplicity
Variable
FLEX selections | Technical excellence | Simplicity |
---|---|---|
Fundamental selection Meet your basic measurement needs |
Technical excellence
|
Simplicity
|
Lean selection Handle your core processes easily |
Technical excellence
|
Simplicity
|
Extended selection Optimize your processes with innovative technologies |
Technical excellence
|
Simplicity
|
Xpert selection Master your most challenging applications |
Technical excellence
|
Simplicity
Variable |
-
Measuring principle
Differential pressure
-
Characteristic
Electronic differential pressure transmitter with ceramic sensor (Ceraphire) for level, volume or mass measurement in liquids.
-
Supply voltage
4...20 mA HART:
12...45V DC (Non Ex)
Ex ia: 12...30V DC -
Reference Accuracy
0.075% of individual sensor,
"PLATINUM" 0.05% of individual sensor -
Long term stability
0.05% of URL/year of individual sensor
-
Process temperature
–25...+150°C
(–13...+302°F) -
Ambient temperature
–40...+80°C
(–40...+176°F) -
Measuring cell
100 mbar...40 bar
(1.5 psi...600 psi) -
Vacuum resistance
0 mbar
-
Max. Turn down
100 : 1
-
Max. overpressure limit
60 bar (900 psi)
-
Process connection
Threads
Flansch (DIN, ASME, JIS) -
Process connection hygienic
DIN11851
DIN11864-1
Tri-Clamp
DRD
Varivent -
Material process membrane
Ceramic
316L, AlloyC -
Material gasket
Viton, Kalrez, EPDM, NBR, Silicone
-
Fill fluid
Silicone Oil
-
Material housing
Die-cast aluminum
Stainless steel -
Communication
4...20 mA HART
-
Certificates / Approvals
ATEX, FM, CSA, IECEx, NEPSI, INMETRO, UK Ex
-
Design approvals
NACE MR0175,
EN10204-3.1, -
Hygienic approvals
EHEDG,
3A
-
Measuring principle
Differential pressure
-
Characteristic / Application
Electronic differential pressure transmitter with ceramic sensor (Ceraphire) for level, volume or mass measurement in liquids.
-
Supply / Communication
4...20 mA HART:
12...45V DC
Exia: 12...30V DC -
Accuracy
0.075% of individual sensor,
"PLATINUM" 0.05% of individual sensor -
Long term stability
0.05% of URL/year of individual sensor
-
Ambient temperature
–40...+80°C
(–40...+176°F) -
Process temperature
–25...+150°C
(–13...+302°F) -
Process pressure / max. overpressure limit
60 bar (900 psi)
-
Pressure measuring range
100mbar...40bar
(1.5psi...600psi) -
Process connection
Threads
Flanges (DIN, ASME, JIS) -
Process connection hygienic
DIN11851
DIN11864-1
Tri-Clamp
DRD
Varivent -
Communication
4...20 mA HART
-
Certificates / Approvals
ATEX, FM, CSA, CSA C/US, IEC Ex, NEPSI, INMETRO
-
Design approvals
NACE MR0175
EN10204-3.1 -
Hygienic approvals
FDA
3A -
Options
4-line digital display
SS- or Aluminium housing -
Application limits
Use Software Applicator Sizing Electronic DP
-
-
Operating Instruction (BA)
Deltabar FMD71, FMD72 Operating Instructions
English version - 02/2018
New version available in English
Electronic differential pressure transmitter for level measurement
EN 26/02/201808/06/201514/04/201415/11/201301/08/2012Language:
English
Version:
26/02/2018
File size:
3.4 MB
File name:
BA01044PEN_0617.pdf
-
-
-
Technical Information (TI)
TI Deltabar FMD71, FMD72
English version - 11/2024
New version available in English
Level measurement with electronic differential pressure HART
EN 18/11/202401/04/202406/10/202126/02/201808/06/2015Language:
English
Version:
18/11/2024
File size:
2.3 MB
File name:
TI01033PEN_0924-00.pdf
-
-
-
Technical Information (TI)
TI Weld-in adapter, process adapter and flanges
-
Technical Information (TI)
English version - 12/2024Level, pressure and temperature measurement
EN 06/12/202425/03/202415/09/202328/10/202130/07/2021Language:
English
Version:
06/12/2024
File size:
1.3 MB
File name:
TI00426FEN_2624.pdf
-
-
-
Technical Information (TI)
TI Deltabar FMD71, FMD72
-
Technical Information (TI)
English version - 11/2024Level measurement with electronic differential pressure
HARTEN 18/11/202401/04/202406/10/202126/02/201808/06/2015Language:
English
Version:
18/11/2024
File size:
2.3 MB
File name:
TI01033PEN_0924-00.pdf
-
-
-
Operating Instruction (BA)
Deltabar FMD71, FMD72 Operating Instructions
-
Operating Instruction (BA)
English version - 02/2018Electronic differential pressure transmitter for level measurement
EN 26/02/201808/06/201514/04/201415/11/201301/08/2012Language:
English
Version:
26/02/2018
File size:
3.4 MB
File name:
BA01044PEN_0617.pdf
-
-
-
Device Parameter (GP)
Deltabar FMD71, FMD72 Description of Device Parameters
-
Device Parameter (GP)
English version - 02/2018Electronic differential pressure transmitter for level measurement
EN 26/02/201808/06/201514/04/201415/11/201301/08/2012Language:
English
Version:
26/02/2018
File size:
711.1 KB
File name:
GP01013PEN_0617.pdf
-
-
-
Short Instruction (KA)
Conversion instructions Cerabar S PMC/PMP7x and Deltabar S PMD/FMD7x from software version 1.x to software Version 2.x or higher
-
Short Instruction (KA)
English version - 05/2021Conversion instructions Cerabar S PMC/PMP7x and Deltabar S
PMD/FMD7x from software version 1.x to software Version 2.x or higherEN 19/05/2021Language:
English
Version:
19/05/2021
File size:
2.3 MB
File name:
KA00615PA2_0108.pdf
-
-
-
Short Instruction (KA)
Deltabar FMD71, FMD72 Brief Operating Instructions
-
Short Instruction (KA)
English version - 02/2018Electronic differential pressure transmitter for level measurement
EN 26/02/201808/06/201514/04/201423/10/201320/07/2012Language:
English
Version:
26/02/2018
File size:
2 MB
File name:
KA01105PEN_0617.pdf
-
-
-
Short Instruction (KA)
Cerabar S, Deltabar S, Deltapilot S - Exchange of the push buttons
-
Short Instruction (KA)
English version - 05/2021Cerabar S, Deltabar S, Deltapilot S - Exchange of the push buttons
EN 17/05/2021Language:
English
Version:
17/05/2021
File size:
396.6 KB
File name:
KA00610PA2_0312.pdf
-
-
-
Installation Instructions (EA)
Installation instruction - Ground terminal T14/T15/T17, F30/F31
-
Installation Instructions (EA)
English version - 06/2015Cerabar S, Deltabar S, Deltabar, Cerabar M, Deltabar M, Deltapilot M
EN 22/06/2015Language:
English
Version:
22/06/2015
File size:
282.8 KB
File name:
EA01131PA2_0115.pdf
-
-
-
Installation Instructions (EA)
Plug M12 Installation instructions
-
Installation Instructions (EA)
English version - 08/2016Deltabar
EN 05/08/2016Language:
English
Version:
05/08/2016
File size:
406.7 KB
File name:
EA01175PA2_0116.pdf
-
-
-
Installation Instructions (EA)
Screw set
-
Installation Instructions (EA)
English version - 03/2020Cerabar PMC51B, PMC71B
Cerabar M PMC51, Cerabar S PMC71
Deltabar FMD71EN 31/03/2020Language:
English
Version:
31/03/2020
File size:
129.1 KB
File name:
EA01296PEN_0120.pdf
-
Choose from our suggested options
-
-- selected coding-- selected label99.999,00<%label%>
-
Looking for more options? Please use the individual configurator to continue
-
-- selected coding-- selected label99.999,00<%feature%>
-
<%label%>99.999,00
-
Looking for more options? Please use the individual configurator to continue
-
Process Connection
-
<%label%>99.999,00
-
<%label%>99.999,00
-
-
<%feature%><%coding%><%label%>
- CAD drawings CAD drawings
- CAD drawings
-
Prices per piece
-
<%range%><%price%>
-
- Optional add-ons
Individual configuration
Often bought together
Accessories
We value your privacy
We use cookies to enhance your browsing experience, collect statistics to optimize site functionality, and deliver tailored advertisements or content.
By selecting "Accept all", you consent to our use of cookies.
For further details please review our cookie policy .