Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Product picture SS2000e single channel H2O, CO2, gas analyzer, front view

SS2000E

SS2000E
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
from CHF ??.-
Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    TDLAS

  • Product Headline

    Single channel gas analyzer for moisture (H2O) in the natural gas industry. The sensor measures gas using tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the gas without coming into physical contact with the stream. Includes a heated enclosure for the sample conditioning system.

  • Channels

    1

  • Analyte and Measurement ranges

    H2O (Moisture):
    0 to 100 ppmv (minimum)
    0 to 5000 ppmv (maximum)

  • Measured Variables

    Concentration
    Moisture Dew Point
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 700 to 1400 mbara or 700 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    100 to 240 VAC, 50 to 60 HZ
    or
    18 to 24 VDC - optional
    1 Amp maximum at 120 VAC, 1.6 Amps at 24 VDC (unheated)
    2 Amps maximum at 120 VAC (heated)

  • Communication

    Analog: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load
    Serial: RS232C - standard, RS485 and Ethernet - optional
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 2, General Fault and Concentration/Assignable Alarm

  • Housing materials

    Electronics: 304 or 316L stainless steel
    Sample System Enclosure: 4X 304 or 316L stainless steel

  • Hazardous area approvals

    CSA Class I, Division 2
    CSA Class I, Zone 2

  • Degree of protection

    Type 4X

  • Measuring principle

    TDLAS

  • Product Headline

    Single channel gas analyzer for carbon dioxide (CO2) in the natural gas industry. The sensor measures gas using tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the gas without coming into physical contact with the stream. Includes a heated enclosure for the sample conditioning system.

  • Channels

    1

  • Analyte and Measurement ranges

    CO2 (Carbon Dioxide):
    0 to 5% volume (minimum)
    0 to 50% volume (maximum)

  • Measured Variables

    Concentration
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 700 to 1400 mbara or 700 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    100 to 240 VAC, 50 to 60 HZ
    or
    18 to 24 VDC - optional
    1 Amp maximum at 120 VAC , 1.6 Amps at 24 VDC (unheated)
    2 Amps maximum at 120 VAC (heated)

  • Communication

    Analog: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load
    Serial: RS232C - standard, RS485 and Ethernet - optional
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 2, General Fault and Concentration/Assignable Alarm

  • Housing materials

    Electronics: 304 or 316L stainless steel
    Sample System Enclosure: 4X 304 or 316L stainless steel

  • Hazardous area approvals

    CSA Class I, Division 2
    CSA Class I, Zone 2

  • Degree of protection

    Type 4X