Try searching by key words, order- or product code or serial number e.g. “CM442” or “Technical information”
Enter at least 2 characters to start the search.
Memosens CCS120D: Digital total chlorine sensor

Digital total chlorine sensor
Memosens CCS120D

Memosens disinfection sensor for wastewater and process water

CCS120D
Please Wait (spinning wheel)
Loading price
Prices temporarily unavailable
Price on request
activating price
price: provide information
from CHF ??.-
Filter Measuring / Application task
Measuring / Application task

Configure to download the CAD drawings of this product

  • Measuring principle

    Total chlorine

  • Application

    Wastewater treatment plants, disinfection monitoring and control in wastewater outlet, reuse of outlet water, utilities of all industries, all disinfection measuring points

  • Characteristic

    Amperometric measurement of total chlorine

  • Measurement range

    0.1 to 10 mg/l total chlorine

  • Measuring method

    Total chlorine consists of free chlorine (HOCl, OCl-) and bound chlorine (chloramines)
    All components are reduced at the working electrode
    Almost pH independent

  • Design

    Closed, membrane covered measuring cell

  • Material

    Sensor shaft: PVC
    Membrane: PET
    Membrane cap: PPE

  • Dimension

    Diameter: 25 mm (0.98 inch)
    Length: 148 mm (5.83 inch)

  • Process temperature

    5 to 45 °C
    (41 to 113 °F)

  • Process pressure

    1 to 4 bar (14.5 to 58 psi)

  • Temperature sensor

    10k NTC integrated (Memosens)

  • Connection

    Inductive, digital connection head with Memosens